::: UNITECH INSTRUMENTATION :::
MUMBAI - 400 086

DATA SCANNER

* Maximum 8 analogue input with any combination.
* Universal input type for thermocouple J, K , R, S, T, RTD
   (PT- 100), 4 - 20mA, 0 - 10V, please specify input while order.
* Automatic cold junction compensation for thermocouples.
* Automatic wire length compensation for long wire RTD inputs.
* Very high accuracy 0.25%, With software linearization.
* Temperature sensor break, ‘oPEn’ indication.
* Under Range ‘undr’& over range ‘OuEr’ indication for 4 to 20 mA &
   0 - 10VDC linear input.
* Easy on-line field calibration by user by front keys.
* Built in temperature reading in °C or F by user selectable.
* Individual channel SKIP, HOLD facility from front panel.
* Reading offset facility to compensate the field error.
* Field programmable channel scan time 1 to 59 seconds.
* Two independent set points individual for each channels with two
   common relay.
* Hysterisis : Programmable for relay output.
* Output - 1 NO/C/NC Relay contact rated 230 V/5Amp. for both relay
   1 (ALARM 1) & relay 2 ( ALARM 2) common for all channels.
* Programmable latch & non-latch facility for common relay &
   relay Indication.
* Optional RS-232 / RS-485 serial port with modbus RTU protocol.
* Optional on line Data Acquisition computer software.
* Optional Parallel port for direct printout on 80 column dotmatrix /
   printer without need of computer(PC).
* Field programmable print interval 5 to 9999 seconds. Printers
   with real time clock for on line printing.
* Optional individual relay output can be given for each channel
 

 The TC-800 multiple data scanner is a part of TC series intelligent Microcontroller based Indicator
 instrument for realization of economic solution of process control problem.
 It is mostly used in:
 * Pharmaceutical validation * Online recording for Oven & BOD, Incubator, Humidity Chamber.
 * Winding temperature of generators, bearings, motors, transformers
 * Heat treatment in furnace. *Water pumping station, water purification & treatment plant.
 * Primary & auxiliary temperature scanning of boiler.
 * Iron & steel industries for soaking pits oxygen plate, smelter plant.
 * Sugar industry for Pan alternation, dailer automation.
 * Chemical & petrochemical industry.

No. of I/P channel: Maximum upto 8 Channels in any Input combination as per users”requirement.

Compensation :

Automatic Cold Junction Compensation for all thermocouples.
Display : 4 digits of 13 mm. height L.E.D for process value display
1 digit of 13mm. height L.E.D for channel number display.
8 L.E.Ds for Alarm Indication. AL1 L.E.D get ON, AL 2 L.E.D's get continous blink.
Setting : 5 feathers touch keys on front panel.
Alarm : 2 independent set points individual for each channels with 2 common
relay output.
Output : 1 NO/C/NC Relay contact rated 230V/5A for relay1 (AL1) & relay2 (Al2).
Port : RS-232/RS-485 Serial port with Modbus output or parallel centronics port for 80 column computer printer for direct printout.
Supply : 230VAC A10%, OR 110VAC A10%, 50Hz.
Mounting : Panel mounting.
Dimension : 96(W) X 96(H) X 175(D) in mm.
Cutout : 92(W) X 92(H) in mm.
Cabinet : ABS plastic, white colour.
Optional : a) Datalog Data Acquisition windows based software for real time trending..
B ) Parallel port for direct printout on 80 column dot-matrix / printer without
need of computer(PC)
C) Individual relay O/P can be given for each channel or with grouping O/P.
Weight : 750 gram (approx.).

 
 

TC-1600 : CLASSIC FEATURES:-
 * Maximum 16 analogue input with any combination.
 * Universal input type for thermocouple J, K , R, S, T,
   RTD (PT- 100), 4 - 20mA, 0 - 10V.(user can connect
   any input)
 * Automatic cold junction compensation for thermocouples.
 * Automatic wirelength compensation for long wire RTD
    inputs.
 * Very high accuracy 0.25%, With software linearization.
 * Temperature sensor break, ‘oPEn’ indication.
 * Under Range ‘undr’& over range ‘OuEr’indication for
    4 to 20 mA & 0 - 10VDC linear input.
 * Easy on-line field calibration by user by front keys.
 * Individual channel SKIP, HOLD facility from front panel.
 * Reading offset facility to compensate the field error.
 * Field programmable channel scan time 1 to 59 seconds.
 * Two independent set points individual for each channels
    with four common relay output with grouping facility.
 * Hysterisis : Programmable for relay output.
 * 1 NO/C/NC Relay contact rated 230 V/5Amp. for four
    relays.
 * Programmable latch & non-latch facility for common
    relay & relay Indication.
 * Optional RS-232 / RS-485 serial port with modbus RTU
    protocol.
 * Optional on line Data Acquisition computer software.
 * Optional Parallel port for direct printout on 80 column
     dotmatrix /  printer without need of computer(PC).
 * Field programmable print interval 1 to 59 minute. Printers
    with real time clock for on line printing.
 * Optional individual relay output can be given for each
    channel.
 

 No. of I/P channels: Maximum upto 16 Channels in any Input combination as per users requirement.

Compensation : Automatic Cold Junction Compensation for all thermocouples
Display : 5 digits of 13 mm. height L.E.D for process value display.
2 digit of 13mm. height L.E.D for channel number display.
32 L.E.Ds for Alarm Indication. AL1 & AL 2.
8 L.E.Ds for TX, RX, HOLD, PROG., Common relay output status.
Setting : 8 feathers touch keys on front panel.
Alarm : 2 independent set points individual for each channels with 2 common
relay output.
Output : 1 NO/C/NC Relay contact rated 230V/5A for relay1 (AL1) & relay2 (Al2).
Port : RS-232/RS-485 Serial port with Modbus output or parallel centronics port for 80 column computer printer for direct printout.
Supply : 230VAC A10%, OR 110VAC A10%, 50Hz.
Mounting : Panel mounting.
Dimension : 96(W) X 192(H) X 350(D) in mm. , 192(W) X 96(H) X 350(D) in mm.
Cutout : 92(W) X 186(H) in mm. , 186(W) X 92(H) in mm.
Cabinet : M.S. Powder coated panel mounting type.
Optional : a) Datalog Data Acquisition windows based software for real time Trending..
b) Parallel port for direct printout on 80 column dot-matrix / printer without need of computer(PC)
C) Individual relay O/P can be given for each channel or with grouping O/P.
Weight : 3.5KG gram (approx.)